| PDC-8 | PDC-8LED |
|---|---|
| UV Light Source | Hg-Xe Lamp / LED |
| 温度范围 | Room temperature ~150°C when using PDC* *When this attachment is not used, it follows the standard DSC specifications. / Room temperature ~150°C when using PDC* *When this attachment is not used, it follows the standard DSC specifications. |
| Wavelength Range | 240–550 nm / N/A |
| Wavelength Selection | 254, 313, 365, 405, 436 nm / 365, 385, 395, 405 nm |
| Maximum Irradiation Intensity | 500 mW/cm² / 150 mW/cm² |
| Irradiation Control | Can be linked to temperature programs / Can be linked to temperature programs |
| DSC200 | DSC600 |
| Heat Flow Measurement Method | Heat flux with centric heat flow / Heat Flow Measurement Method: Heat flux with centric heat flow and 3 thermocouples |
| Time Constant (resolution) | 6.5 seconds or better / 3 seconds or better |
| RMS noise / Sensitivity | 0.1 µW / 0.2 µW / 0.05 µW / 0.1 µW |
| 温度范围 | -150 to 725°C |
| Programmable Heating/Cooling Rate | 0.01 ~ 100˚C/min |
| Temperature Precision | /- 0.01°C |
| Temperature Accuracy | /- 0.05°C |
| Calorimetric Precision | /- 0.05% |
| Calorimetric Accuracy | /- 0.4% |
| Baseline Stability | /-5µW |
| Real View® Camera System | Optional |