规格参数。Industry Application example Pharmaceuticals, medical technology & biotechnology TOC monitoring in WFI production systems, TOC control in cleaning validation, TOC measu
广西科米瑞实验仪器设备有限公司生产销售 HORIBA Tocadero EVO,可提供产品参数、报价和售后服务咨询,价格面议。
← 返回产品列表
Tocadero EVO
| Industry | Application example |
|---|---|
| Pharmaceuticals, medical technology & biotechnology | TOC monitoring in WFI production systems, TOC control in cleaning validation, TOC measurement in cooling water circuits, and TOC monitoring in pharmaceutical wastewater. |
| Semiconductor & Electronics manufacturing | TOC monitoring in ultrapure water systems, TOC control in process chemicals, TOC measurement in CMP slurries, and TOC monitoring in semiconductor wastewater. |
| Seawater desalination | TOC measurement in seawater inflow, TOC control in membrane cleaning solutions, and TOC monitoring in concentrate effluent. |
| Hydrogen production | TOC control in electrolysis water, TOC monitoring in cooling circuits, TOC measurement in electrolyte solutions, and TOC monitoring in process wastewater. |
| Energy generation/ Power plants | TOC monitoring in boiler feed water, TOC control in cooling water systems, TOC measurement in condensates, and TOC monitoring in power plant wastewater. |
| Drinking water treatment | TOC measurement in raw water, TOC monitoring after treatment stages, TOC control in pure water, and TOC monitoring in municipal wastewater. |
| Cosmetics & Personal Care Industry | TOC control in production water, TOC monitoring in cooling systems, TOC measurement in process solutions, and TOC monitoring in cosmetic wastewater. |
| Chemical industry | TOC monitoring in reactor cooling water, TOC control in process water, TOC measurement in catalyst suspensions, and TOC monitoring in chemical wastewater. |
| Methods & Detection | Continuous flow with UV oxidation and conductivity detection |
| Cycle Time | 1 sec. (online method) 1 min. + 1 min. rinsing (offline method) |
| Sample Flow | approx. 14 ml/min |
| Measurement Range | 0 - 1000 ppb TOC Application-dependent up to 2000 ppb TOC |
| Accuracy & Precision | Limit of detection: 0,1 ppb Accuracy: ± 0,05 ppb online |
| Connections | 1 x sample, 1 x check 1 x sample, 3 x check (optional) |
| Normalizations | USP <643>, EP 2.2.44, JP, CP FDA 21 CFR Part 11 - Compliance for Pharma |
| Environmental Conditions | Relative humidity: < 80 % non condensing Temperature: 10 - 40 °C |
| Sample Type | Pure Water (PW) & Ultra Pure Water (UPW), Water for Injection (WFI), Boiler Feed Water |
| Conductivity Sample | 2 μS/cm Application-dependent up to < 10 μS/cm |
| Input Pressure | < 0,5 bar |
| Sample Temperature | < 50 °C |
| Interfaces | 2 x analog (0/4 - 20 mA), according NAMUR NE43 1 x relais (collective error) RJ45 Ethernet, USBOptional: 2 additional analog outputs, 2 additional relay outputs, autosampler |
| Dimensions | 338 W x 200 D x 450 H mm |
| Weight | 11 kg |
| Power | 110 - 230 VAC ± 10 %, 47 - 63 Hz, 60 Watt |
请填写必填信息,我们将在工作日尽快回复。