MO1000 COD/BOD
U.S. and Canada Technical Representatives Research and Development Partnerships
广西科米瑞实验仪器设备有限公司提供 MANTECH MO1000 COD/BOD 产品资料、选型咨询、供货报价和售后服务支持,价格面议。
← 返回产品列表
U.S. and Canada Technical Representatives Research and Development Partnerships PeCOD® COD/BOD Analyzer Online COD & BOD Analyzer – MO1000 PeCOD® Optimized TOC Analyzer Online TOC Analyzer – MO1000 MT Automated Environmental Titration System MANTECH Pro Software BOD Pro Biochemical Oxygen Demand Analyzer BOD Software – Benchtop Solution Software Packages BOD Software – Benchtop Solution MANTECH Pro Software Industrial & Municipal Wastewater Treatment Analysis
| 10mL Total Volume – Single Analysis | 20mL Total Volume – Duplicate and Triplicate Analysis |
|---|---|
| Anions | Formula / Remark |
| Ammonium | NH4+ / No interference for NH4+ ≤ 500 ppm Note: Similar results are obtained for Ammonia |
| Carbonate | CO32- / No interference for CO32- ≤ 500ppm using chloride resistant sensor |
| Chlorate | ClO3– / No interference for ClO3– ≤ 500ppm |
| Nitrate | NO3– / No interference, NO3– can be used as PeCOD electrolyte |
| Nitrite | NO2– / No interference for NO2– ≤ 500ppm |
| Perchlorate | ClO4– / No interference, ClO4– can be used as PeCOD electrolyte |
| Phosphate | PO43- / No interference for PO43- ≤ 500ppm |
| Sulfate | SO42- / No interference for SO42- ≤ 500ppm |
| Sulfite | SO32- / Interference for SO3 ≥ 20 ppm, giving COD high by 90% at 250 ppm SO32- |
| Sulfide | S2- / Interference for S2- > 0 ppm, giving COD high by >100% at 50 ppm S2- |
| Cations | Formula / Remark |
| Aluminum | Al3+ / No interference for Al3+ ≤ 500ppm |
| Calcium | Ca2+ / No interference for Ca2+ ≤ 500ppm |
| Chromate | Cr3+ / Interference for Cr3+ > 2 ppm, giving low COD |
请填写必填信息,我们将在工作日尽快回复。