Series 2
All instruments include a 110V AC adapter with US prong style. Select alternative prongs from the Accessories for all Series 2 Gauges table.
广西科米瑞实验仪器设备有限公司生产销售 Mark-10 Series 2,可提供产品参数、报价和售后服务咨询,价格面议。
← 返回产品列表All instruments include a 110V AC adapter with US prong style. Select alternative prongs from the Accessories for all Series 2 Gauges table.
©2026 Mark-10 Corporation , All Rights Reserved
Mark-10 defines accuracy as a percentage of full scale of the instrument. To determine the measurement error as an actual load value, multiply the accuracy percentage by the instrument’s capacity.
The accuracy is ±0.1% of full scale (FS). Multiply ±0.1% by 50 lbF, which equals ±0.05 lbF. This means that any displayed reading may be higher or lower by up to 0.05 lbF. For example, if the displayed value is 30.00 lbF, the true reading will be ≥29.95 lbF and ≤30.05 lbF.
The accuracies of the sensor and the indicator must be added together. Models 7i and 5i indicators have accuracy values of ±0.1% FS, while the Model 3i is rated at ±0.2% FS. Using the example of a Series R50 torque sensor with Model 3i indicator, add ±0.35% to ±0.2%, which equals ±0.55%. In a specific example for the Model MR50-12, the accuracy becomes ±0.55% x 135 Ncm = ±0.7425 Ncm.
请填写必填信息,我们将在工作日尽快回复。