Tablet Disintegration Tester
This test is provided to determine whether tablets or capsules disintegrate within the prescribed time when placed in a liquid medium at certain experimental conditions.
广西科米瑞实验仪器设备有限公司生产销售 Koehler Instrument Tablet Disintegration Tester,可提供产品参数、报价和售后服务咨询,价格面议。
← 返回产品列表
This test is provided to determine whether tablets or capsules disintegrate within the prescribed time when placed in a liquid medium at certain experimental conditions. Meets current specifications of USP, IP, and other Pharmacopoeias Specifically designed for use in the quality and production control of normal plain coated tablets and gelatin capsules Programmable Temperature and Time Molded, Clear acrylic water bath with illumination for bettervisibility Baskets are automatically lifted at the end of the test Capability for dual buffer disintegration test Audible, low water level alarm with indication on display for safety. External temperature sensors available for individual beaker The Tablet Disintegration Tester is advance microcontroller based with user friendly software in dialog mode. The Alpha Numeric splash proof keyboard allows entries such as samplename, sample number or ID number for data authentication.The report printout complies with GLP requirements and abuilt-in Real Time Clock provides date, time display, and authenticity of the printout. Operating Modes: Count or Time, user selectable Dip Speed: 30 ± 1 DPM Stroke Length: 5.5 ± 0.1 cm Water Bath Temperature Range: ambient +5 to 50°C Water Bath Temperature Accuracy: ± 0.2° Water Bath Circulation: Immersion Pump Beaker Volume: 1000 mL Basket: 2 x 6 test positions (10 mesh) USP Type A