SIM970
10 MΩ input impedance External trigger for data sync Unique continuous autocalibration 90 dB power line frequency rejection True 5½-digit performance
广西科米瑞实验仪器设备有限公司提供 Stanford Research Systems SIM970 产品资料、选型咨询、供货报价和售后服务支持,价格面议。
← 返回产品列表
10 MΩ input impedance External trigger for data sync Unique continuous autocalibration 90 dB power line frequency rejection True 5½-digit performance Four isolated channels Bright 7-segment LED displays Three decade autoranging Continuous autocalibration 90 dB line frequency rejection SIM970 5½-Digit Voltmeter (4 Ch.) The SIM970 Quad Digital Voltmeter is designed to make precision DC voltage measurements with excellent long-term accuracy . For applications in which many voltages must be monitored, up to 16 DVM channels can be put into one SIM900 mainframe. Four voltage ranges from ±199.999 mV to ±19.9999 V can be autoranged or manually selected. Programming Examples 3D Step File SIM Modules Scientific Instruments SIM Modules Sci. Instruments SIM970 5½-Digit Voltmeter The SIM970 Quad Digital Voltmeter is designed to make precision DC voltage measurements with excellent long-term accuracy . For applications in which many voltages must be monitored, up to 16 DVM channels can be put into one SIM900 mainframe. Four voltage ranges from ±199.999 mV to ±19.9999 V can be autoranged or manually selected. An external trigger input allows synchronization of voltage readings on all four channels for critical applications requiring coincidental readings. A BUSY output gives a TTL (logic high) when readings are being taken. Autocalibration is performed with every reading by sequentially measuring not only the input voltage, but also the ground and the full-scale voltages against a calibrated internal reference. This autocalibration routine virtually eliminates offsets and scale errors, and ensures smooth range-to-range transitions . The bright front-panel LED display shows updated readings three times per second. Computer access through the SIM900 mainframe ( RS-232 or GPIB) permits data logging with 24 bits of resolution. All channels are isolated from ground, and from each other. The SIM970 uses isolated BNC connectors for inputs so coaxial cables can be used for reduced noise pickup. SIM970 Specifications 5½ (±199999 counts) [ 1 ] ±19.9999 V, 100 µV res., 1.5 counts noise (rms) [ 1, 2 ] ±1.99999 V, 10 µV res., 0.8 counts noise (rms) [ 1, 2 ] ±999.99 mV, 10 µV res., 0.8 counts noise (rms) [ 1, 2 ] ±199.999 mV, 1 µV res., 1.0 counts noise (rms) [ 1, 2 ] ±(% of reading + counts) [ 3 ] 0.0010 + 2 (24 hr.), 0.0050 + 2 (90 day), 0.0080 + 2 (1 yr.) 0.0002 + 2 (24 hr.), 0.0050 + 2 (90 day), 0.0080 + 2 (1 yr.) 0.0002 + 4 (24 hr.), 0.0050 + 6 (90 day), 0.0080 + 6 (1 yr.) (24 hour counts error)/2 [ 3 ][ 5 ] (typ.) 10 MΩ ± 1 %, >3 GΩ selectable on ranges 2 to 4 [ 6 ] BNC (Amphenol 31-10 or similar) ±60 V center to shield, ±200 V shield to earth Internal, external (TTL), or remote
请填写必填信息,我们将在工作日尽快回复。
| 技术参数 | 10 MΩ input impedance |
|---|---|
| 技术参数 | 90 dB power line frequency rejection |
| 技术参数 | Bright 7-segment LED displays |
| 技术参数 | 90 dB line frequency rejection |
| 技术参数 | The SIM970 Quad Digital Voltmeter is designed to make precision DC voltage measurements with excellent long-term accuracy . For applications in which many voltages must be monitored, up to 16 DVM channels can be put into one SIM900 mainframe. Four vo |
| 技术参数 | The SIM970 Quad Digital Voltmeter is designed to make precision DC voltage measurements with excellent long-term accuracy . |
| 技术参数 | For applications in which many voltages must be monitored, up to 16 DVM channels can be put into one SIM900 mainframe. Four voltage ranges from ±199.999 mV to ±19.9999 V can be autoranged or manually selected. An external trigger input allows synchro |
| 技术参数 | Autocalibration is performed with every reading by sequentially measuring not only the input voltage, but also the ground and the full-scale voltages against a calibrated internal reference. This autocalibration routine virtually eliminates offsets a |
| 技术参数 | The bright front-panel LED display shows updated readings three times per second. Computer access through the SIM900 mainframe ( RS-232 or GPIB) permits data logging with 24 bits of resolution. All channels are isolated from ground, and from each oth |
| 技术参数 | 10 MΩ ± 1 %, >3 GΩ selectable on ranges 2 to 4 [ 6 ] |
| 技术参数 | 1 s to within 3 counts of final reading on ranges 1 to 3, 8 s on range 4 |
| 技术参数 | Red LED, 0.40", with polarity indication. Green LED for range and autorange indication |