For the detection of hydrogen sulfide in liquefied petroleum (LP) gases. The sensitivity of the test is about 4 mg/m3 (0.15 to 0.2 grain of hydrogen sulfide per 100 ft3) of gas.
广西科米瑞实验仪器设备有限公司生产销售 Koehler Instrument Hydrogen Sulfide Apparatus,可提供产品参数、报价和售后服务咨询,价格面议。
← 返回产品列表
For the detection of hydrogen sulfide in liquefied petroleum (LP) gases. The sensitivity of the test is about 4 mg/m3 (0.15 to 0.2 grain of hydrogen sulfide per 100 ft3) of gas. Conforms to the specifications of ASTM D2420 Vaporized LP gas is passed over moist lead acetate paper under controlled conditions. Hydrogen sulfide reacts with lead acetate to form lead sulfide which produces a coloration on the paper varying from yellow to black, depending upon the amount of hydrogen sulfide present. Liquefied petroleum gases and their products of combustion must not be unduly corrosive to the materials with which they come in contact. The potential personnel exposure hazards of H 2 S also make the detection and measurement of hydrogen sulfide important, even in low concentrations. In addition, in some cases the odor of the gases must not be objectionable. KD2420 Glassware Kit Comprised of: Glass Cylinder Rubber Stopper (2) Watch Glass Glass Rod / Hook K27853 LPG Sampling Cylinder, 500mL Conforms to the specifications of: ASTM D2420, ISO 8819 Temperature Range: +5 to 150°C Temperature Control Stability: ± 0.05°C Temperature Readout Accuracy: ± 0.5°C Bath Capacity: 28L (7.3gal) LPG Cylinder Capacity: 500mL Flow Meter Range: 2 – 3 L/min