Particle characterization instrumentBrookhaven Instruments型号 BI-DCP – Disc Centrifuge Photosedimentometer
The Brookhaven BI-DCP Disc Centrifuge Photosedimentometer (DCP) is a digitally controlled, high-resolution, particle size distribution analyzer. It yields high-resolution result...
- Analysis
- Size range of 0.01 to 30 µm maximum; 0.05 to 10 µm typical for low density particles, 0.01 to 2 µm for high density particles; wide variety of materials dispersed in water and other solvents
- Data System
- Windows computer with printer. Real time graphics and data display, high resolution color (SVGA). Menu driven data analysis and management system, with single copy license and software distribution copy.
- Drive System
- Microprocessor controlled digitally driven electronic motor. Digital readout for setting and monitoring speed. Speed continuously variable from 500 to 15,000 rpm. Speed accuracy and stability better than + 0.01%. Temperature sensor and digital readou
- Disc Cavity
- Polymethylmethacrylate with stainless steel hub. Dynamically balanced over range of rotational speeds. Spin fluid volume from 10 to 40 ml.